Main Features:
- Lower cost than DTGS/DLATGS.
- Lower N.E.P. than DTGS/DLATGS.
- Faster frequency response than DTGS/DLATGS.
- Wider spectral response than windowed DTGS/DLATGS.
- Higher operating temperature range than DTGS/DLATGS.
SKU: -
Sciencetech's new proprietary wideband detector design comes packaged
in our standard "hockey puck" enclosure (2" diameter x 1" high) and is
connectorized to be powered directly from the Stanford SR-810/830 type
lock-in amplifiers and is optimized to utilize the full input range of
these lock-ins for FTIR applications.
This detector is not hygroscopic and may be operated indefinitely under
normal atmospheric conditions and at temperatures beyond 100 degrees C.
A selection of windows is available to limit spectral response.
Our current specifications: (subject to change without notice)
| Model Number: |
SCICV400-B0 (Opamp type) |
| Active sensor size: |
2mm. dia. |
| Spectral response: |
Flat 100nm. - 1mm. (no window, blackend) |
| Frequency response: |
Flat 0.25Hz - 750Hz (blackened) |
| Responsivity: |
10000 V/W (10Hz typical, 100nm-1mm, no window, blackend) |
| Output saturation: |
+/- 2 volts (Stanford overload lights before saturation) |
| Output offset: |
+/- 5 mV (typical, 25 deg C) |
| Input saturation: |
200x10-6 Watts |
| Damage threshold: |
1x106 Watts |
| N.E.P.: |
4.05x10-10 Watts (4uV out, typical, 10Hz, 1Hz bandwidth) |
| D* (at 10Hz): |
4.37x108 cm*Hz.5/W |
| Dynamic range: |
2.5x105:1 (18 bits) |
Price: $ 3,250.00 USD
Price: Please request price at aquaglia@sciencetech-inc.com
Sciencetech is working in the development of a linear array of these detectors. The first intended array will have only 16 detectors.
For availability of the above detectors, please contact Alex Quaglia:
Email: aquaglia@sciencetech-inc.com
Phone: (519) 644-0135 ext. 109
Fax: (519) 644-0136
1450 Global Dr.,
London, ON.
Canada, N6N 1R3
Price: $3,250.00