Sciencetech TeraHertz Lithium Tantalate (LiTaO3)

Pyroelectric Detector Head Listing
|
Sciencetech TeraHertz Lithium Tantalate Pyroelectric Detector Heads
|
Note: Our pyroelectric detectors have inherently flat response from 100nm-1000μm but are window limited.
Note2: Pyroelectric detectors REQUIRE chopped radiation input and a synchronous output detection system.
Note3: Pyroelectric detectors may be sensitive to rapid temperature change, vibration and/or cell phones.
|
| Model | Sensor | Window | Sensitivity* |
NEP* | SatP* | Cal ? | Options |
$(no options) |
| SCISDX1011-options | 5mm Dia. | Quartz | 420-480 V/W |
9e-9 | 10mW | Yes 6328 | W C L X |
$1907 |
| SCIEL404CM-options | 5mm Dia. | Quartz | 160-240 V/W |
2e-7 | 20mW | No | W C L X |
$ 669 |
| SCIEL420CM-options | 2.7mm Sq. | Quartz | 80-120kV/W |
2e-8 | 30μW | No | W C L X |
$ 323 |
| SCIEL420BJ-options | 2.7mm Sq. | None | 0.8-1.2MV/J |
5e-8J | 3μJ | No | W C L X |
$ 509 |
| Options | Description | $ Additional |
| No option selected |
Standard. As listed above with separate 6ft output and power cables.
Power connector is DB9 female with Sciencetech universal pinout for
connection to SCIPWRMOD power and measuring system. Compatible
with EOS PS-1 power supply or EOS LIA/PS-1 lock-in/power supply.
Package: 2" dia. x 1" long with 1" dia. face mount and 1/4-20 side mount.
Power requirements are +/- 8 volts to +/-24 volts at +/- 8mA max.
Quartz window allows essentially flat response (+/- 3%) 300nm-2.1μm.
Electrical response -3dB points typically 0.5Hz - 150Hz.
|
$ 0.00 |
| W-#option Window | -A Sapphire, -G Germanium, -S Silicon, -C CaF2, -B BaF2, -K KBr, -0 None |
$ 208 |
| C-#option Calibrate |
Calibrate at specified wavelength(s): . . . . (+/-1% traceable 632.8nm)
+/-5% for each other selected wavelength . . . 200nm - 2.5μm (+/-2%)
-All = 100nm intervals from 300nm to 2100μm with +/-5% accuracy.
|
$ 243
$ 448
$ 1894 |
| L Lockin module |
Piggyback synchronous detector board is added internally to head.
Accepts external reference OR performs homodyne detection on signal
equal to or greater than 0.1% of full scale. Unity gain, TC=0.1/1.0sec.
Additional power requirement : +/- 8 volts to +/-24 volts at +/- 5mA max.
BNC reference connector + switches on rear of 1" longer packaging.
|
$ 485 |
| X Experimental |
Any customer specified gain, response, connection or packaging.
|
Contact us |
| Additional notes and clarifications: |
| Sensitivity* |
Measured (and calibrated) in a 1 Hz bandwidth at 22 degrees Celsius using a 632.8nm
source chopped at 8.5 Hz with 50/50% duty cycle.
Warning: Response (and calibration value) will be reduced by 3% with 50/50 chopped
input at 2 octaves (4x/25%) above/below the -3dB points. i.e. -3% at 2Hz and 40Hz typ.
On board jumper converts sensitivity from total power to power density. (uncalibrated)
|
| NEP* |
Typical, units are Watts/Hertz^.5, measured as per sensitivity.
|
| SatP* |
Saturation power. i.e. Output just below clip level.
|
| Quartz |
May be quartz or fused silica.
|
| For broadband UNCHOPPED applications, check out Sciencetech's new fully compatible THERMOPILE head.
|
|