Sciencetech Lithium Tantalate Pyroelectric Detectors


Sciencetech TeraHertz Lithium Tantalate Pyroelectric Detector Heads
Note: Our pyroelectric detectors have inherently flat response from 100nm-1000μm but are window limited.
Note2: Pyroelectric detectors REQUIRE chopped radiation input and a synchronous output detection system.
Note3: Pyroelectric detectors may be sensitive to rapid temperature change, vibration and/or cell phones.
Model Sensor Window Sensitivity* NEP* SatP* Cal ? Options $(no options)
SCISDX1011-options 5mm Dia. Quartz 420-480 V/W 9e-9 10mW Yes 6328 W C L X $1907
SCIEL404CM-options 5mm Dia. Quartz 160-240 V/W 2e-7 20mW No W C L X $ 669
SCIEL420CM-options 2.7mm Sq. Quartz 80-120kV/W 2e-8 30μW No W C L X $ 323
SCIEL420BJ-options 2.7mm Sq. None 0.8-1.2MV/J 5e-8J 3μJ No W C L X $ 509
Options Description $ Additional
No option selected Standard. As listed above with separate 6ft output and power cables.
Power connector is DB9 female with Sciencetech universal pinout for
connection to SCIPWRMOD power and measuring system. Compatible
with EOS PS-1 power supply or EOS LIA/PS-1 lock-in/power supply.
Package: 2" dia. x 1" long with 1" dia. face mount and 1/4-20 side mount.
Power requirements are +/- 8 volts to +/-24 volts at +/- 8mA max.
Quartz window allows essentially flat response (+/- 3%) 300nm-2.1μm.
Electrical response -3dB points typically 0.5Hz - 150Hz.
$ 0.00
W-#option Window -A Sapphire, -G Germanium, -S Silicon, -C CaF2, -B BaF2, -K KBr, -0 None $ 208
C-#option Calibrate Calibrate at specified wavelength(s): . . . . (+/-1% traceable 632.8nm)
+/-5% for each other selected wavelength . . . 200nm - 2.5μm (+/-2%)
-All = 100nm intervals from 300nm to 2100μm with +/-5% accuracy.
$ 243
$ 448
$ 1894
L Lockin module Piggyback synchronous detector board is added internally to head.
Accepts external reference OR performs homodyne detection on signal
equal to or greater than 0.1% of full scale. Unity gain, TC=0.1/1.0sec.
Additional power requirement : +/- 8 volts to +/-24 volts at +/- 5mA max.
BNC reference connector + switches on rear of 1" longer packaging.
$ 485
X Experimental Any customer specified gain, response, connection or packaging. Contact us
Additional notes and clarifications:
Sensitivity* Measured (and calibrated) in a 1 Hz bandwidth at 22 degrees Celsius using a 632.8nm
source chopped at 8.5 Hz with 50/50% duty cycle.
Warning: Response (and calibration value) will be reduced by 3% with 50/50 chopped
input at 2 octaves (4x/25%) above/below the -3dB points. i.e. -3% at 2Hz and 40Hz typ.
On board jumper converts sensitivity from total power to power density. (uncalibrated)
NEP* Typical, units are Watts/Hertz^.5, measured as per sensitivity.
SatP* Saturation power. i.e. Output just below clip level.
Quartz May be quartz or fused silica.
For broadband UNCHOPPED applications, check out Sciencetech's new fully compatible THERMOPILE head.